Standard design




Portable ophthalmic scan SW1000P, ophthalmic a scan, ophthalmic scanner,ophthalmic equipment

ophthalmic a scan      SW1000P

With macular recognition function, can accurately measure axial length, anterior chamber depth, lens thickness, as well as intra-ocular lens calculation.

Composed of main unit, probe, and so on.
The probe contacts the cornea and sends out ultrasound of 20MHz. The
echo waves from front and back surface of cornea are detected. The signal
is processed and then we get the thickness of the cornea. The data are
displayed on the color LCD screen through output interface, or stored in the
machine, or printed out.


Characteristics and parameter
P scan probe: 20MHZ, angle of 45 degrees makes easier operation;
P scan precision: +-0.01mm;
Measuring range: 0.3mm~1.2mm;
Each ID can save 10 groups of data, and each group is the average of
20 measures;
ID could be input and easy to check;
Output: Internal thermal printer;
Indication: Big size chromatic LCD;
Input: Touch screen;
Gain adjusting: 0-127db;
Power Supply: AC 220 V±10%,50 Hz±2%;
Power: 30W+15%;
Specification of protective tube:F1AL250Vφ4×12mm;

Test results comply with GB9706.1-1995 Class I Type B and
GB9706.9.